The precision and agreement of corneal thickness and keratometry measurements with SS-OCT versus Scheimpflug imaging

Abstract

Purpose

To assess the repeatability and reproducibility of swept-source optical coherence tomography (SS-OCT) and Scheimpflug system and evaluate the agreement between the two systems in measuring multiple corneal regions in children.

Methods

Pachymetric and keratometric maps for both systems were evaluated. Central, midperipheral and peripheral corneal thickness (CT), keratometry and astigmatism power vectors were recorded. The three outcomes yielded by the same observer were used to assess intraobserver repeatability. The differences in the mean values provided by each observer were used to evaluate interobserver reproducibility. Within-subject standard deviation, test-retest repeatability (TRT) and coefficient of variation (CoV) were used to analyze the intraobserver repeatability and interobserver reproducibility. Paired T-test and Bland-Altman were used to appraise interdevice agreement.

Results

Seventy-eight eyes of 78 children were included. The CoV was ≤2.12 and 1.10%, respectively, for repeatability and reproducibility. TRT and CoV were lower for central and paracentral CT measurements than for peripheral measurements. The SS-OCT device generated higher precision when acquiring CT data, whereas Scheimpflug system showed higher reliability when measuring corneal keratometry. Although the CT readings measured using SS-OCT were significantly thinner than Scheimpflug device (P <  0.001), the central and thinnest CT values were still of high agreement. The interdevice agreement of keratometry measurement was high for the central corneal region and moderate for the paracentral and peripheral areas.

Conclusions

The precision of CT measurements by SS-OCT was higher, while the reliability of keratometry measurements by the Scheimpflug system was higher in children. Apart from the measured values in the central corneal region, the thickness and keratometry readings should not be considered interchangeable between the two systems.

Background

Precise measurement of corneal thickness (CT) and refractive power in children is vital for screening corneal ectasia, monitoring myopia progression, and planning orthokeratology [1, 2]. The measurements are important as it not only includes the central cornea, but also the peripheral zone, and alterations in these could indicate the development of corneal diseases, such as keratoconus and Fuchs’ endothelial dystrophy [3, 4].

To obtain a topographic map of the cornea, various technologies including Placido disk corneal topography, slit-scanning corneal topography, Scheimpflug imaging and optical coherence tomography (OCT) have been employed. Placido disk imaging does not provide information regarding the posterior corneal surface. Slit-scanning generates a lower repeatability in characterizing the posterior corneal surface when compared with the Scheimpflug principle [5].

Several reports have revealed high precision of rotating Scheimpflug camera, the Pentacam (Oculus Optikgeräte GmbH, Wetzlar, Germany), and an anterior-segment OCT (AS-OCT), the CASIA SS-1000 (Tomey, Nagoya, Japan) in measuring the central corneal thickness (CCT) and power [5,6,7,8,9]. OCT is considered a high-resolution, real-time ocular imaging technology. Time-domain OCT combined with Placido disk corneal topographer [10,11,12,13], spectral-domain OCT (SD-OCT) with or without Placido disk imaging [14,15,16], and swept-source OCT (SS-OCT) were commercially released for acquiring the topographic map of cornea [17, 18]. Several studies have reported high precision of anterior segment SS-OCT, CASIA (SS-1000; Tomey, Nagoya, Japan), in acquiring pachymetric and keratometric data of the central cornea [17,18,19]. However, there is no study till date that has investigated the precision of these devices in measuring peripheral cornea under similar conditions. Additionally, there are no published papers that measured corneal topography in children, and its extent of cooperation remained low, challenging the reliability of measurement.

Thus, the purpose of this study was to comprehensively assess the intraobserver repeatability and interobserver reproducibility of the above-mentioned Scheimpflug camera and SS-OCT as well as to evaluate the interdevice agreement when measuring multiple corneal regions in children with myopia.

Methods

Subjects

This prospective study was conducted at the Eye Hospital of Wenzhou Medical University. The research protocol adhered to the tenets of the Declaration of Helsinki, and was approved by the Office of Research Ethics, Wenzhou Medical University (KYK2013–21). Signed informed consent forms by the guardians of subjects were obtained before undergoing examinations.

The exclusion criteria included children with trauma, acute ocular inflammation, any history of contact lens wear, previous ophthalmological surgeries, and ocular diseases other than ametropia. Before being enrolled in this study, all subjects underwent a complete ophthalmic examination, including subjective refraction, ophthalmoscopy, noncontact tonometry (TX-F; Cannon, Tokyo, Japan), slit-lamp microscopy and fundoscopy.

Instruments

CASIA is an anterior segment SS-OCT device that uses a 1310 nm light source and produces a scan range with 6.0 mm depth and 16.0 mm diameter, yielding an axial resolution of ≤10 μm and a lateral resolution of ≤30 μm. The “Corneal Map” mode takes 0.3 s to obtain 16 radial B-scans at a range of 10 mm centered on the apical cornea, and each B-scan comprises of 512 A-scans. The captured information was then processed to generate the topographic map of the cornea.

The Pentacam HR is a high-resolution imaging system that works on the principle of Scheimpflug. It uses a slit-light source operating in a monochromatic blue light at a wavelength of 475 nm, and a 1.45-megapixel Scheimpflug camera rotating on the visual axis for taking 25 or 50 cross-section pictures of the anterior segment. In 2 s, up to 138,000 true elevation points are acquired to construct the corneal topography. The 25-picture scan mode was used in this study.

Measurement procedures

In order to promote children’s compliance, one observer provided detailed instructions to each subject before beginning the measurement, and additionally the parent demonstrated to the child on how to cooperate during the examination. The subject was seated in a dim room with the chin on the chinrest and forehead against the forehead bar and was asked to fixate on the specified fixation point with both eyes wide open. Each device was manipulated according to the user’s manual. The scanning by Pentacam HR was automatically initiated when the corneal vertex was centered and focused manually, whereas the CASIA measurement was triggered manually after the alignment procedure was automatically accomplished by the system. To assure the measurement independence, patients were asked to move their head away from the chinrest, and the scan units were thoroughly retreated before subsequent examinations. The Pentacam data were considered valid if the “QS” index of the measurement showed “OK”. As for CASIA, the B-scan images were reviewed by the two observers individually after each measurement, ensuring that there was no apparent image artifact for OCT images. In addition, the observers also carefully performed checks for corneal maps to verify the scan quality.

Three successive scans were performed by each observer (DC and HFZ) between 9 AM and 5 PM. The observers were trained to use the device 1 month before this study began. The sequence of the 2 devices and the 2 examiners were randomly set. The time of whole measurement process for each subject was rigorously controlled within 20 min. The three measurements with each system were used to assess intraobserver repeatability. The outcomes of the 3 consecutive measurements obtained by the same observer were averaged, and the differences between the observers were used to evaluate interobserver reproducibility. The disparities regarding the parameters measured using CASIA and Pentacam HR were used to appraise the interdevice agreement.

Parameters were recorded on the following three zones:

  1. 1.

    The central zone of the cornea: CT values measured by each device included central corneal thickness (CCT) and thinnest corneal thickness (TCT); corneal power indices including the mean keratometry (Km) along with the steepest and the flattest anterior corneal meridians, and the magnitude and axis of astigmatism were analyzed using power vectors method (J0 and J45) as described by Thibos et al. [20]

  2. 2.

    The paracentral zone of cornea: The CT and keratometry at 2 mm diameter in the nasal (CT2mm-Nasal, K2mm-Nasal), superior (CT2mm-Superior, K2mm-Superior), temporal (CT2mm-Temporal, K2mm-Temporal) and inferior (CT2mm-Inferior, K2mm-Inferior) regions centered on the corneal vertex (Fig. 1).

  3. 3.

    The peripheral zone of cornea: The CT and keratometry at 5 mm diameter in the nasal (CT5mm-Nasal, K5mm-Nasal), superior (CT5mm-Superior, K5mm-Superior), temporal (CT5mm-Temporal, K5mm-Temporal) and inferior (CT5mm-Inferior, K5mm-Inferior) regions centered on the corneal vertex (Fig. 1).

Fig. 1
figure 1

Paracentral and peripheral corneal indices measurements: Point A at 2-mm diameter circle nasally, Point B 2-mm superiorly, Point C 2-mm temporally, Point D 2-mm inferiorly, Point E at 5-mm diameter circle nasally, Point F 5-mm superiorly, Point G 5-mm temporally and Point H 5-mm inferiorly

Statistical analysis

Statistical analyses were performed using SPSS (version 21.0, SPSS, IBM® Co, Armonk, New York, USA) and Microsoft Office Excel (Microsoft® Co, Redmond, Washington, D.C., USA). All data distributions were verified for normality by the Kolmogorov-Smirnov test. The results were expressed as mean ± standard deviation (SD). A P value of less than 0.05 was considered to be statistically significant.

To assess the intraobserver repeatability, one-way analysis of variance (one-way ANOVA) was performed for 3 consecutive measurements by each observer. Within-subject standard deviation (Sw), test-retest repeatability (TRT), within-subject coefficient of variation (CoV), and intraclass correlation coefficients (ICC) were computed. Since astigmatism power vectors have small magnitudes (which make CoV quite large, so that it cannot represent the real variance among measurements) and can be either positive or negative, we did not rely on CoV to estimate their repeatability. Therefore, the precision for measurement of astigmatism power vector were assessed using ICC only. The TRT was calculated as 2.77 × Sw, which was the expected upper limit for 95% of the difference between measurements [21]. The CoV was defined as 100% × Sw / overall means. An ICC higher than 0.9 was considered as high consistency, and an ICC between 0.75 to 0.90 was considered as moderately consistent, and an ICC less than 0.75 was considered as poor consistency [22]. To evaluate the interobserver reproducibility, the mean values obtained by the same observer were calculated, and Sw, TRT, CoV, and ICC were computed for the 2 mean values obtained by the two observers. To appraise the interdevice agreement, paired T-test and Bland-Altman plots were applied, and 95% limits of agreement (LoA) was calculated as mean ± 1.96 SD of the differences between the two instruments.

Results

A total of 78 right eyes from 78 children (47 males and 31 females) diagnosed with refractive errors were recruited. Among them, the proportion for 4 years old was 2.56%, for 5 years old was 3.85%, for 6 years old was 6.41%, for 7 years old was 16.67%, for 8 years old was 20.51%, for 9 years old was 19.23%, for 10 years old was 14.10%, for 11 years old was 10.26%, for 12 years old was 5.13%, and for 14 years old was 1.28%. The mean age of the subjects was 8.55 ± 1.97 years (range, 4–14 years), and the mean spherical equivalent refraction was − 1.49 ± 1.79 D (range, − 5.75 to 2.25 D).

Intraobserver repeatability of corneal thickness measurements

For intraobserver repeatability of CT measurements by both devices, TRT and CoV were lower for CCT, TCT and paracentral CT measurements than for peripheral measurements. When taken individually, the TRT acquired with CASIA ranged from 2.98 to 12.42 μm, and the CoV was lower than 0.75% (Table 1). Both TRT and CoV were relatively higher for CT measurements with CASIA at superior locations of paracentral and peripheral cornea. Pentacam HR showed higher TRT and CoV on a scale of 14.68 to 34.19 μm and 0.98 to 2.12%, respectively (Table 2). The relatively greater TRT and CoV were also noticed when measuring the paracentral and peripheral cornea at inferior location. Comparison of the two instruments demonstrated that the TRT and CoV for CCT and TCT measurements generated by CASIA were lower than a quarter with respect to those provided by Pentacam HR. Regarding the paracentral and peripheral CT measurements, most of the TRTs and CoVs yielded by CASIA were about one-third of those rendered by Pentacam HR.

Table 1 Intraobserver repeatability outcomes for corneal thickness obtained using CASIA swept-source optical coherence tomography in children
Table 2 Intraobserver repeatability outcomes for corneal thickness obtained using Pentacam Scheimpflug imaging in children

Intraobserver repeatability of corneal power measurements

With regards to the intraobserver repeatability of keratometry, the TRT and CoV in the central region were smaller than those in the paracentral and peripheral regions, and among these, the TRT and CoV for K5mm-superior measurements remained the highest. The TRT for CASIA ranged from 0.28 to 1.32 D, and CoVs were ≤ 1.10% (Table 3). The TRT and CoV for Pentacam HR were on a scale of 0.25 to 0.98 D and 0.21 to 0.81%, respectively (Table 4). In addition, the TRT and CoV for central keratometry measurements were comparable between the two systems but were slightly greater with CASIA than with Pentacam HR for the paracentral and peripheral regions.

Table 3 Intraobserver repeatability outcomes for corneal power obtained using CASIA swept-source optical coherence tomography in children
Table 4 Intraobserver repeatability outcomes for corneal power obtained using Pentacam Scheimpflug imaging in children

As shown in Tables 3 and 4, the J0 measurement showed high repeatability for both CASIA and Pentacam HR with ICCs > 0.9. However, the repeatability of J45 measurement remained poor for CASIA with an ICC <  0.75, and moderate for Pentacam HR with an ICC ranging from 0.766 to 0.832.

Interobserver reproducibility of corneal thickness measurements

With regards to the interobserver reproducibility of CT measurements, the TRT and CoV were higher for the paracentral and peripheral zones than those for CCT and TCT measurements. The TRT and CoV generated by CASIA were lower than 7.85 μm and 0.48%, respectively (Supp Table 1), while Pentacam HR yielded higher TRT and CoV ranging from 10.01 to 21.08 μm and from 0.67 to 1.31%, respectively (Supp Table 2). In comparison, the CoVs for all locations with CASIA were approximately one-third of those obtained with Pentacam HR.

Interobserver reproducibility of corneal power measurements

With regards to the interobserver reproducibility of keratometry, the TRT and CoV for central cornea measurement were smaller when compared to those for paracentral and peripheral areas, and the highest TRT and CoV were observed for K5mm-superior measurements. The TRT and CoV for cornea measurements with CASIA ranged from 0.25 to 0.72 D and 0.21 to 0.60%, respectively (Supp Table 3). For Pentacam HR, the TRT and CoV were ≤ 0.41 D and 0.34%, respectively (Supp Table 4). Comparison showed that TRT and CoV at all locations rendered by CASIA were higher than those provided by Pentacam HR.

The ICC for J0 measurement with CASIA was > 0.9, but for J45 was only 0.784 (Supp Table 3). As for Pentacam HR, the ICC value obtained was > 0.9 for both J0 and J45 measurements (Supp Table 4).

Interdevice agreement of corneal thickness measurements

Table 5 shows significantly thinner CT measurements with CASIA than with Pentacam HR (P <  0.0001 in all cases, paired T-test). The width of 95% LoA was the smallest for CCT and TCT measurements, and greater for peripheral CT acquirements (Fig. 2 and Fig. 3). For both the paracentral and peripheral CT measurements, relatively wider 95% LoAs were observed for the superior location.

Table 5 The difference and agreement for corneal thickness measurements between CASIA swept-source optical coherence tomography and Pentacam Scheimpflug imaging in children
Fig. 2
figure 2

Bland–Altman plots showing pair-wise agreement between CASIA and Pentacam HR for central corneal thickness

Fig. 3
figure 3

Bland–Altman plots showing pair-wise agreement between CASIA and Pentacam HR for superior corneal thickness at 5 mm diameter circle centered on the corneal vertex

Interdevice agreement of corneal power measurements

Table 6 shows the differences in keratometry measurements between CASIA and Pentacam HR (Fig. 4 and Fig. 5). The 95% LoA was narrowed for keratometry measurement in the central and temporal regions. Furthermore, the agreement of keratometry measurement for paracentral cornea was lower than that for the peripheral CT measurement.

Table 6 The difference and agreement for corneal power measurements between CASIA swept-source optical coherence tomography and Pentacam Scheimpflug imaging in children
Fig. 4
figure 4

Bland–Altman plots showing pair-wise agreement between CASIA and Pentacam HR for mean of keratometry metrics (Km) along the steepest and the flattest anterior corneal meridians

Fig. 5
figure 5

Bland–Altman plots showing pair-wise agreement between CASIA and Pentacam HR for superior corneal keratometry at 5 mm diameter circle centered on the corneal vertex

Discussion

In this study, SS-OCT and rotating Scheimpflug camera were comprehensively used to assess the repeatability and reproducibility of CT and refractive power measurements in multiple regions of the cornea in children. The precision (repeatability and reproducibility) evaluation is a mandatory task to test the reliability of new technology, and some findings have been reported in adults previously. Szalai et al. [18] noticed a better repeatability of CCT measurements with CASIA (TRT = 4.17 μm) than with the Pentacam HR (TRT = 7.33 μm), and these values were similar to our study outcomes. The same study reported a contradictory finding wherein CASIA showed a lower repeatability (TRT = 21.922 μm) than Pentacam HR (TRT = 11.451 μm) when measuring TCT. Moreover, the repeatability for keratometry measurements with CASIA (TRT, 0.481–0.555 D) and Pentacam HR (TRT, 0.468–0.472 D) were both lower than that in our study. Neri et al. [23] noticed a better repeatability for CCT measurement using CASIA than that using a spectral-domain OCT (Cirrus-OCT, Carl Zeiss Meditec AG, Germany). Though they listed SD only, we acquired a CoV of 0.33% by calculating the ratio of SD to the corresponding mean value. This value was close to our study results.

Our study showed that the two systems had high precision when measuring both thickness and keratometry metrics in central cornea, with a slight declination towards the periphery. High repeatability for CT and corneal curvature measurements have been reported [7, 24,25,26,27]. Xu et al. [28] used Pentacam HR to measure CT and also found a downward trend for precision from central to peripheral region. As explicated before [7], due to the distribution of scan lines in a radial pattern around the visual axis, more points were captured for the central cornea analysis when compared with those at the peripheral region. It was worth noting that the peripheral superior corneal measurements generated the worst precision among all locations with SS-OCT, which was in agreement with the results of our previous study using a spectral-domain OCT to measure CT [7]. The upper eyelashes possibly covered the cornea at this location, and may thus lead to deterioration on the precision of measurement. A similar outcome was found for K measurement using Pentacam HR but the peripheral inferior CT measurement showed the lowest precision. Not surprisingly, the visible blue light that the Pentacam used along with the relatively long scan time could prevent some children from keeping their eyes wide open, making both upper and lower cornea measurement susceptible to the interference of eyelids or eyelashes. However, an unexpected finding was that, apart from the superior positions, the precision for paracentral corneal keratometry measurements with Pentacam HR was lower than that for the peripheral area. A likely explanation for this outcome might be due to the imperfect reconstruction algorithm of Pentacam HR for keratometry measurements in this zone. To the best of our knowledge, this study is the first to evaluate paracentral corneal (the nasal, superior, temporal and inferior position at a distance of 1 mm to the corneal vertex) keratometry.

The precision of astigmatism power vector measurements remained poor with CASIA (repeatability ICC for J0, 0.930 to 0.933; ICC for J45, 0.715 to 0.724) when compared to Pentacam HR (repeatability ICC for J0, 0.961 to 0.962; ICC for J45, 0.766 to 0.832). Previous studies [29, 30] reported similar results for Pentacam HR when measuring J0 (ICC, 0.974 to 0.979) and J45 (ICC, 0.876 to 0.888). These attributed the moderate precision to the small value of corneal astigmatism, and the same explanation could be used for our study. The absolute value of J0 derived from the children was greater than that of J45, and the measurement of J0 showed higher ICC value than J45 measurement. Savini et al. [14] used AS-OCT combined with Placido corneal topography (MS-39, Costruzione Strumenti Oftalmici, Florence, Italy) to measure the total corneal astigmatism, and found better repeatability of both J0 (ICC = 0.975) and J45 (ICC = 0.950) measurements. Taken together, it seems that the Placido disk could improve the astigmatism measurement.

An interesting finding observed in our study was that the SS-OCT outperformed the Scheimpflug-based corneal topographic map in measuring CT, while Scheimpflug-based corneal topographer was observed to be more precise for corneal power measurement. Firstly, the values of repeatable and reproducible CoVs for CT measurements were smaller with CASIA when compared with Pentacam HR. Our previous study compared RTVue and Pentacam in obtaining CT measurements at the same locations as set currently, and the results showed a better repeatability with SD-OCT than with Pentacam [7]. We considered that this disparity might be due to high resolution and short acquisition time of OCT technology. It should be noted that the high-resolution version of Pentacam was employed this time, but higher CoVs for CT measurements were discovered when compared with those reported by our study previously (CoVs, 0.98–2.12% vs. 0.65–1.10%) [7]. A likely reason for this may be related to the lower cooperative degree of children. Despite this, CASIA still generated a slightly higher repeatability than RTVue (CoVs, 0.20–0.75% vs. 0.31–1.16%) when measuring CT [7]. This was probably because the automatic alignment function applied to CASIA minimized the impact resulting from off-axis measurement.

Secondly, the CoV was marginally greater for corneal power measurement with CASIA at each location, indicating better precision with Pentacam HR when acquiring keratometric map of the cornea. This discrepancy was also reported for CASIA and Pentacam HR for the measurements of anterior keratometry on the central cornea [18, 19]. Wang et al. [16] found low precision of keratometry measurement using another SD-OCT (RTVue; repeatable ICC, 0.982–0.990; reproducible TRT, 0.26–0.44 D, reproducible CoV, 0.22–0.36%), as compared with the measurement using CASIA in the current study (repeatable ICC, 0.994–0.995; reproducible TRT, 0.26 D, reproducible CoV, 0.21%). Savini et al. [14] employed SD-OCT combined with Placido device (MS-39) and revealed a higher repeatability of keratometry measurement with a TRT of 0.20 D and a CoV of 0.16%. The simple OCT instrument had a limited role in measuring cornea power, and the combination of OCT and Placido-disk imaging is considered to be an effective means for improving the accuracy. Several studies [11,12,13] also generated high repeatability of TD-OCT when combined with Placido instrument (Omni, Carl Zeiss Meditec AG, Germany) in measuring corneal power indices.

Though the CT readings provided by CASIA were significantly thinner than those measured by Pentacam HR, the interdevice agreement still remained high for the central cornea. However, the agreement was from moderate to poor for CT measurements from the paracentral to the peripheral region. Milla et al. [26] compared a TD-OCT (Visante) and a Scheimpflug-Placido corneal topographer (Sirius, Costruzione Strumenti Oftalmici, Florence, Italy) when measuring the CCT and peripheral CT. These have set the peripheral CT measurements on a distance of 2.5 mm and 4.0 mm, respectively, to the corneal vertex, and showed smaller CT readings at all locations with OCT instrument, as well as poor agreement between Visante and Sirius with 95% LoA of − 42.7 to − 2.0 μm for CCT measurement, and − 42.8 to 24.0 μm, − 59.1 to 7.6 μm, − 77.5 to 9.2 μm, and − 51.6 to 0.4 μm, respectively for 2.5 mm temporally, nasally, superiorly, and inferiorly, and − 86.1 to − 32.9 μm, − 112.2 to − 22.8 μm, − 88.7 to 15.2 μm, and − 128.9 to 6.2 μm, respectively, for 4.0 mm temporally, nasally, superiorly, and inferiorly. OCT instruments tended to underestimate the CCT values as compared to Scheimpflug-based devices in the normal cornea [11, 26, 31,32,33,34,35,36,37], and SS-OCT is not an exception [17,18,19, 38,39,40,41].

With regards to the corneal power measurements, high interdevice agreement was observed for the central cornea. Nakagawa et al. [17] reported lower agreement between the two instruments (95% LoA, − 1.00 to 1.90 D) when measuring the central corneal power, as compared with the current result. A similar outcome has been reported by Szalai et al. [18], which was consistent with our study result. Ghoreishi et al. [38] noticed a high agreement between CASIA and Pentacam HR (95% LoA, − 0.24 to 0.54 D) in adults. As compared with the peripheral regions, agreement for paracentral keratometry measurements was even lower, which resulted from the abnormal precision of Pentacam HR for keratometry measurements in this area.

A limitation of our study would be that only one model of OCT was used. Further investigation is warranted to compare more OCT instruments, including TD-OCT and SD-OCT combined with Placido disk devices. Additionally, investigations are required to determine the precision when enrolling children with abnormal corneas, such as congenital corneal opacities, macrocornea and microcornea.

Conclusion

In summary, both CASIA SS-OCT and Pentacam high-resolution Scheimpflug system showed high precision when measuring CT and keratometry in children, although a slight decrease in precision was noted for the peripheral cornea. Furthermore, the reliability of CT measurement was higher with the SS-OCT device, while the precision of corneal power measurement was higher with the Scheimpflug imaging system. Therefore, we recommend the use of a pachymetric map of the cornea acquired with AS-OCT and a corneal keratometric map obtained with rotating Scheimpflug camera in clinical practice. In addition, the interdevice agreement of CT measurement was high for the central cornea zone, but moderate for the paracentral and peripheral regions. With respect to measuring corneal power, high agreement was observed when measuring by keratometry in central regions. Hence, only the central and the TCT as well as keratometry in the central area can be used interchangeably between the two devices.

Availability of data and materials

All data generated or analyzed during this study are included in this published article.

Abbreviations

OCT:

Optical coherence tomography

SS-OCT:

Swept-source optical coherence tomography

AS-OCT:

Anterior-segment optical coherence tomography

SD-OCT:

Spectral-domain OCT

K:

Keratometry

K2mm-Nasal, K2mm-Superior, K2mm-Temporal, K2mm-Inferior :

The keratometry at 2 mm diameter in the nasal, superior, temporal and inferior regions centered on the corneal vertex.

K5mm-Nasal, K5mm-Superior, K5mm-Temporal, K5mm-Inferior :

The keratometry at 5 mm diameter in the nasal, superior, temporal and inferior regions centered on the corneal vertex.

CT:

corneal thickness

CT2mm-Nasal, CT2mm-Superior, CT2mm-Temporal, CT2mm-Inferior :

The corneal thickness at 2 mm diameter in the nasal, superior, temporal and inferior regions centered on the corneal vertex.

CT5mm-Nasal, CT5mm-Superior, CT5mm-Temporal, CT5mm-Inferior :

The corneal thickness at 5 mm diameter in the nasal, superior, temporal and inferior regions centered on the corneal vertex.

CCT:

Central corneal thickness

TCT:

Thinnest corneal thickness

Mean ± SD:

Mean ± standard deviation

TRT:

Test-retest variability

CoV:

Coefficient of variation

ICC:

Intra-class correlation coefficient

Sw :

Intra-subject standard deviation

LoA:

Limit of agreement

References

  1. Zhong Y, Chen Z, Xue F, Zhou J, Niu L, Zhou X. Corneal power change is predictive of myopia progression in orthokeratology. Optom Vis Sci. 2014;91(4):404–11.

    Article  PubMed  Google Scholar 

  2. Léoni-Mesplié S, Mortemousque B, Touboul D, Malet F, Praud D, Mesplié N, et al. Scalability and severity of keratoconus in children. Am J Ophthalmol. 2012;154(1):56–62.e1.

    Article  PubMed  Google Scholar 

  3. Brautaset RL, Nilsson M, Miller WL, Leach NE, Tukler JH, Bergmanson JP. Central and peripheral corneal thinning in keratoconus. Cornea. 2013;32(3):257–61.

    Article  PubMed  Google Scholar 

  4. Repp DJ, Hodge DO, Baratz KH, McLaren JW, Patel SV. Fuchs' endothelial corneal dystrophy: subjective grading versus objective grading based on the central-to-peripheral thickness ratio. Ophthalmology. 2013;120(4):687–94.

    Article  PubMed  PubMed Central  Google Scholar 

  5. de Jong T, Sheehan MT, Koopmans SA, Jansonius NM. Posterior corneal shape: comparison of height data from 3 corneal topographers. J Cataract Refract Surg. 2017;43(4):518–24.

    Article  PubMed  Google Scholar 

  6. McAlinden C, Khadka J, Pesudovs K. A comprehensive evaluation of the precision (repeatability and reproducibility) of the Oculus Pentacam HR. Invest Ophthalmol Vis Sci. 2011;52(10):7731–7.

  7. Huang J, Ding X, Savini G, Pan C, Feng Y, Cheng D, et al. A comparison between Scheimpflug imaging and optical coherence tomography in measuring corneal thickness. Ophthalmology. 2013;120(10):1951–8.

    Article  PubMed  Google Scholar 

  8. Savini G, Hoffer K, Carbonelli M, Barboni P. Scheimpflug analysis of corneal power changes after myopic excimer laser surgery. J Cataract Refract Surg. 2013;39(4):605–10.

    Article  PubMed  Google Scholar 

  9. Hernández-Camarena J, Chirinos-Saldaña P, Navas A, Ramirez-Miranda A, de la Mota A, Jimenez-Corona A, et al. Repeatability, reproducibility, and agreement between three different Scheimpflug systems in measuring corneal and anterior segment biometry. J Refract Surg. 2014;30(9):616–21.

    Article  PubMed  Google Scholar 

  10. Srivannaboon S, Chotikavanich S, Chirapapaisan C, Kasemson S, Po-ngam W. Precision analysis of posterior corneal topography measured by Visante Omni: repeatability, reproducibility, and agreement with Orbscan II. J Refract Surg. 2012;28(2):133–8.

    Article  PubMed  Google Scholar 

  11. Viswanathan D, Kumar NL, Males JJ, Graham SL. Comparative analysis of corneal measurements obtained from a Scheimpflug camera and an integrated Placido-optical coherence tomography device in normal and keratoconic eyes. Acta Ophthalmol. 2015;93(6):e488–94.

    Article  PubMed  Google Scholar 

  12. Shah JM, Han D, Htoon HM, Mehta JS. Intraobserver repeatability and interobserver reproducibility of corneal measurements in normal eyes using an optical coherence tomography-Placido disk device. J Cataract Refract Surg. 2015;41(2):372–81.

    Article  PubMed  Google Scholar 

  13. Martínez-Albert N, Esteve-Taboada J, Montés-Micó R. Repeatability of whole-cornea measurements using an anterior segment imaging device based on OCT and Placido-disk. Expert Rev Med Devices. 2017;14(2):169–75.

    Article  PubMed  CAS  Google Scholar 

  14. Savini G, Schiano-Lomoriello D, Hoffer KJ. Repeatability of automatic measurements by a new anterior segment optical coherence tomographer combined with Placido topography and agreement with 2 Scheimpflug cameras. J Cataract Refract Surg. 2018;44(4):471–8.

    Article  PubMed  Google Scholar 

  15. Tang M, Chen A, Li Y, Huang D. Corneal power measurement with Fourier-domain optical coherence tomography. J Cataract Refract Surg. 2010;36(12):2115–22.

    Article  PubMed  PubMed Central  Google Scholar 

  16. Wang Q, Hua Y, Savini G, Chen H, Bao F, Lin S, et al. Corneal power measurement obtained by Fourier-domain optical coherence tomography: repeatability, reproducibility, and comparison with Scheimpflug and automated keratometry measurements. Cornea. 2015;34(10):1266–71.

    Article  PubMed  Google Scholar 

  17. Nakagawa T, Maeda N, Higashiura R, Hori Y, Inoue T, Nishida K. Corneal topographic analysis in patients with keratoconus using 3-dimensional anterior segment optical coherence tomography. J Cataract Refract Surg. 2011;37(10):1871–8.

    Article  PubMed  Google Scholar 

  18. Szalai E, Berta A, Hassan Z, Módis L Jr. Reliability and repeatability of swept-source Fourier-domain optical coherence tomography and Scheimpflug imaging in keratoconus. J Cataract Refract Surg. 2012;38(3):485–94.

    Article  PubMed  Google Scholar 

  19. Szalai E, Nemeth G, Hassan Z, Módis L Jr. Noncontact evaluation of corneal grafts: swept-source Fourier domain OCT versus high-resolution Scheimpflug imaging. Cornea. 2017;36(4):434–9.

    Article  PubMed  Google Scholar 

  20. Thibos LN, Wheeler W, Horner D. Power vectors: an application of Fourier analysis to the description and statistical analysis of refractive error. Optom Vis Sci. 1997;74(6):367–75.

    Article  CAS  PubMed  Google Scholar 

  21. Bland JM, Altman DG. Measurement error. BMJ. 1996;313(7059):744.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  22. Mcgraw KO, Wong SP. Forming inferences about some intraclass correlation coefficients. Psychological Methods. 1996;1(1):30–46.

    Article  Google Scholar 

  23. Neri A, Malori M, Scaroni P, Leaci R, Delfini E, Macaluso C. Corneal thickness mapping by 3D swept-source anterior segment optical coherence tomography. Acta Ophthalmol. 2012;90(6):e452–7.

    Article  PubMed  Google Scholar 

  24. Khoramnia R, Rabsilber T, Auffarth G. Central and peripheral pachymetry measurements according to age using the Pentacam rotating Scheimpflug camera. J Cataract Refract Surg. 2007;33(5):830–6.

    Article  PubMed  Google Scholar 

  25. Bourges JL, Alfonsi N, Laliberté JF, Chagnon M, Renard G, Legeais JM, et al. Average 3-dimensional models for the comparison of Orbscan II and Pentacam pachymetry maps in normal corneas. Ophthalmology. 2009;116:2064–71.

    Article  PubMed  Google Scholar 

  26. Milla M, Piñero D, Amparo F, Alió J. Pachymetric measurements with a new Scheimpflug photography-based system: intraobserver repeatability and agreement with optical coherence tomography pachymetry. J Cataract Refract Surg. 2011;37(2):310–6.

    Article  PubMed  Google Scholar 

  27. Kawamorita T, Nakayama N, Uozato H. Repeatability and reproducibility of corneal curvature measurements using the Pentacam and Keratron topography systems. J Refract Surg. 2009;25(6):539–44.

    Article  PubMed  Google Scholar 

  28. Xu Z, Peng M, Jiang J, Yang C, Zhu W, Lu F, et al. Reliability of Pentacam HR thickness maps of the entire cornea in normal, post-laser in situ keratomileusis, and keratoconus eyes. Am J Ophthalmol. 2016;162:74–82.e1.

  29. Chen D, Lam AK. Reliability and repeatability of the Pentacam on corneal curvatures. Clin Exp Optom. 2009;92(2):110–8.

    Article  PubMed  Google Scholar 

  30. Aramberri J, Araiz L, Garcia A, Illarramendi I, Olmos J, Oyanarte I, et al. Dual versus single Scheimpflug camera for anterior segment analysis: precision and agreement. J Cataract Refract Surg. 2012;38(11):1934–49.

    Article  PubMed  Google Scholar 

  31. Prospero Ponce CM, Rocha KM, Smith SD, Krueger RR. Central and peripheral corneal thickness measured with optical coherence tomography, Scheimpflug imaging, and ultrasound pachymetry in normal, keratoconus-suspect, and post-laser in situ keratomileusis eyes. J Cataract Refract Surg. 2009;35(6):1055–62.

    Article  PubMed  Google Scholar 

  32. Fu J, Wang X, Li S, Wu G, Wang N. Comparative study of anterior segment measurement with Pentacam and anterior segment optical coherence tomography. Can J Ophthalmol. 2010;45(6):627–31.

    Article  PubMed  Google Scholar 

  33. Huang J, Pesudovs K, Yu A, Wright T, Wen D, Li M, et al. A comprehensive comparison of central corneal thickness measurement. Optom Vis Sci. 2011;88(8):940–9.

    Article  PubMed  Google Scholar 

  34. Ishibazawa A, Igarashi S, Hanada K, Nagaoka T, Ishiko S, Ito H, et al. Central corneal thickness measurements with Fourier-domain optical coherence tomography versus ultrasonic pachymetry and rotating Scheimpflug camera. Cornea. 2011;30(6):615–9.

    Article  PubMed  Google Scholar 

  35. Chen S, Huang J, Wen D, Chen W, Huang D, Wang Q. Measurement of central corneal thickness by high-resolution Scheimpflug imaging, Fourier-domain optical coherence tomography and ultrasound pachymetry. Acta Ophthalmol. 2012;90(5):449–55.

    Article  PubMed  Google Scholar 

  36. Randleman JB, Lynn MJ, Perez-Straziota CE, Weissman HM, Kim SW. Comparison of central and peripheral corneal thickness measurements with scanning-slit, Scheimpflug and Fourier-domain ocular coherence tomography. Br J Ophthalmol. 2015;99(9):1176–81.

    Article  PubMed  Google Scholar 

  37. Yazici AT, Bozkurt E, Alagoz C, Alagoz N, Pekel G, Kaya V, et al. Central corneal thickness, anterior chamber depth, and pupil diameter measurements using Visante OCT, Orbscan, and Pentacam. J Refract Surg. 2010;26(2):127–33.

    Article  PubMed  Google Scholar 

  38. Ghoreishi SM, Mortazavi SAA, Abtahi ZA, Abtahi MA, Sonbolestan SA, Abtahi SH, et al. Comparison of Scheimpflug and swept-source anterior segment optical coherence tomography in normal and keratoconus eyes. Int Ophthalmol. 2017;37(4):965–71.

    Article  PubMed  Google Scholar 

  39. Chan TCY, Biswas S, Yu M, Jhanji V. Comparison of corneal measurements in keratoconus using swept-source optical coherence tomography and combined Placido–Scheimpflug imaging. Acta Ophthalmol. 2016;95(6):e486–94.

    Article  PubMed  CAS  Google Scholar 

  40. Lee YW, Choi CY, Yoon GY. Comparison of dual rotating Scheimpflug-Placido, swept-source optical coherence tomography, and Placido-scanning-slit systems. J Cataract Refract Surg. 2015;41(5):1018–29.

    Article  PubMed  Google Scholar 

  41. Fukuda S, Kawana K, Yasuno Y, Oshika T. Anterior ocular biometry using 3-dimensional optical coherence tomography. Ophthalmology. 2009;116(5):882–9.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This work was supported in part by the Medical and Health Science and Technology Program of Zhejiang Province (2019KY111); Foundation of Wenzhou City Science & Technology Bureau (Y20180174); Zhejiang Provincial Key Research and Development Program (2018C03012); Zhejiang Provincial High-level Talents Program (2017–102); Wenzhou Key Team of Scientific and Technological Innovation (C20170002). The contribution from the G.B. Bietti Foundation was supported by the Italian Ministry of Health and Fondazione Roma. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

Author information

Authors and Affiliations

Authors

Contributions

Concept and design YZ, DC, JH, CM; data collection DC, HZ; statistical expertise QW, GS, JH; administrative, technical and logistic support JH, CM; analysis and interpretation YZ, HZ, YJ, BS; writing the article QW, YJ, RN, JH, CM; critical revision of the article YZ, DC, GS, BS, RN, JH; final approval of the article YZ, DC, GS, QW, HZ, YJ, BS, RN, JH, CM. All authors reviewed the manuscript and agreed to be accountable for all aspects of the work.

Corresponding authors

Correspondence to Jinhai Huang or Chenyang Mei.

Ethics declarations

Ethics approval and consent to participate

This study adhered the tenets of the Declaration of Helsinki and was approved by the Research Review Board of the Eye Hospital of Wenzhou Medical University (KYK2013–21). Informed consent was obtained from all patients.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Supplementary information

Additional file 1: Table S1.

Interobserver reproducibility outcomes for corneal thickness obtained using CASIA swept-source optical coherence tomography in children.

Additional file 2: Table S2.

Interobserver reproducibility outcomes for corneal thickness obtained using Pentacam Scheimpflug imaging in children.

Additional file 3: Table S3.

Interobserver reproducibility outcomes for corneal power obtained using CASIA and swept-source optical coherence tomography in children.

Additional file 4: Table S4.

Interobserver reproducibility outcomes for corneal power obtained using Pentacam Scheimpflug imaging in children.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Zhao, Y., Chen, D., Savini, G. et al. The precision and agreement of corneal thickness and keratometry measurements with SS-OCT versus Scheimpflug imaging. Eye and Vis 7, 32 (2020). https://doi.org/10.1186/s40662-020-00197-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40662-020-00197-0

Keywords