Table 1 Intraobserver repeatability of the ZW-30

From: Evaluation of a new dynamic real-time visualization 25 kHz swept-source optical coherence tomography based biometer

Parameter

Observer

Mean ± SD

Sw

TRT

CoV (%)

ICC (95% CI)

AL (mm)

1st

25.76 ± 1.10

0.01

0.02

0.02

1.000 (1.000 to 1.000)

 

2nd

25.79 ± 1.16

0.01

0.03

0.04

1.000 (1.000 to 1.000)

CCT (μm)

1st

543.65 ± 30.34

1.84

5.11

0.34

0.996 (0.995 to 0.998)

 

2nd

537.73 ± 31.39

1.91

5.30

0.36

0.996 (0.994 to 0.998)

AQD (mm)

1st

3.10 ± 0.25

0.01

0.04

0.46

0.997 (0.995 to 0.998)

 

2nd

3.10 ± 0.27

0.02

0.05

0.56

0.996 (0.993 to 0.998)

ACD (mm)

1st

3.64 ± 0.25

0.01

0.04

0.39

0.997 (0.995 to 0.998)

 

2nd

3.64 ± 0.27

0.02

0.05

0.46

0.996 (0.994 to 0.998)

LT (mm)

1st

3.68 ± 0.27

0.02

0.05

0.51

0.995 (0.993 to 0.997)

 

2nd

3.72 ± 0.28

0.02

0.06

0.57

0.994 (0.991 to 0.997)

Km (D)

1st

43.36 ± 1.33

0.12

0.32

0.27

0.992 (0.989 to 0.995)

 

2nd

43.36 ± 1.45

0.11

0.30

0.25

0.995 (0.991 to 0.997)

AST (D)

1st

1.28 ± 0.73

0.12

0.34

0.972 (0.960 to 0.981)

 

2nd

1.10 ± 0.66

0.12

0.34

0.966 (0.946 to 0.980)

J0 (D)

1st

 − 0.59 ± 0.39

0.06

0.18

0.974 (0.963 to 0.982)

 

2nd

 − 0.48 ± 0.37

0.07

0.19

0.967 (0.947 to 0.981)

J45 (D)

1st

 − 0.03 ± 0.18

0.06

0.18

0.887 (0.842 to 0.922)

 

2nd

 − 0.03 ± 0.20

0.07

0.18

0.899 (0.842 to 0.939)

CD (mm)

1st

11.89 ± 0.40

0.07

0.20

0.61

0.967 (0.953 to 0.978)

 

2nd

11.96 ± 0.42

0.08

0.22

0.65

0.967 (0.947 to 0.981)

  1. AL = axial length; CCT = central corneal thickness; AQD = aqueous depth; ACD = anterior chamber depth; LT = lens thickness; Km = mean keratometry; AST = astigmatism; CD = corneal diameter; SD = standard deviation; Sw = within-subject standard deviation; TRT = test–retest repeatability (2.77 Sw); CoV = within-subject coefficient of variation; ICC = intraclass correlation coefficient; CI = confidence interval