Skip to main content

Table 3 Diagnostic efficiency of blinking parameters in DED from Lipiview and K5M

From: Comparison of deep learning-assisted blinking analysis system and Lipiview interferometer in dry eye patients: a cross-sectional study

Parameter

 

Cut-off value

AUC

SE

P value

95% CI

Sensitivity (%)

Specificity (%)

Number of blinks

Lipiview

> 18

0.659

0.065

0.022

(0.532, 0.786)

51.43

74.29

K5M

> 21

0.650

0.067

0.031

(0.518, 0.781)

51.43

80.00

Number of IB

Lipiview

> 9

0.662

0.065

0.020

(0.535, 0.789)

65.71

60.00

K5M

> 8

0.773

0.056

< 0.001

(0.664, 0.883)

74.29

71.43

IB rate

Lipiview

> 0.25

0.531

0.070

0.651

(0.395, 0.668)

88.57

25.71

K5M

> 0.389

0.716

0.063

0.002

(0.593, 0.839)

85.71

57.14

  1. DED = dry eye disease; AUC = area under curve; SE = standard error; CI = confidence interval; K5M = Keratograph 5M; IB = incomplete blinking