Parameter/device | Mean difference ± SD | 95% LoA | P value |
---|---|---|---|
K1 (D) | |||
Anterion vs. Argos | − 0.084 ± 0.294 | − 0.660, 0.493 | 0.176 |
Anterion vs. IOLMaster 700 | − 0.038 ± 0.260 | − 0.548, 0.472 | 0.512 |
Anterion vs. Pentacam AXL | 0.072 ± 0.455 | − 0.821, 0.965 | 0.585 |
Argos vs. IOLMaster 700 | 0.046 ± 0.257 | − 0.458, 0.549 | 0.915 |
Argos vs. Pentacam AXL | 0.156 ± 0.395 | − 0.618, 0.929 | 0.005* |
IOLMaster 700 vs. Pentacam AXL | 0.110 ± 0.446 | − 0.764, 0.984 | 0.041* |
K2 (D) | |||
Anterion vs. Argos | − 0.166 ± 0.341 | − 0.836, 0.504 | 0.001* |
Anterion vs. IOLMaster 700 | − 0.084 ± 0.266 | − 0.606, 0.438 | 0.272 |
Anterion vs. Pentacam AXL | 0.040 ± 0.409 | − 0.762, 0.841 | 0.794 |
Argos vs. IOLMaster 700 | 0.082 ± 0.302 | − 0.509, 0.673 | 0.221 |
Argos vs. Pentacam AXL | 0.206 ± 0.350 | − 0.480, 0.892 | < 0.001* |
IOLMaster 700 vs. Pentacam AXL | 0.124 ± 0.386 | − 0.634, 0.881 | 0.033* |
WTW (mm) | |||
Anterion vs. Argos | − 0.191 ± 0.264 | − 0.709, 0.326 | < 0.001* |
Anterion vs. IOLMaster 700 | − 0.202 ± 0.189 | − 0.574, 0.170 | < 0.001* |
Anterion vs. Pentacam AXL | − 0.111 ± 0.166 | − 0.436, 0.213 | < 0.001* |
Argos vs. IOLMaster 700 | − 0.010 ± 0.262 | − 0.524, 0.503 | 0.168 |
Argos vs. Pentacam AXL | 0.080 ± 0.267 | − 0.444, 0.605 | 0.972 |
IOLMaster 700 vs. Pentacam AXL | 0.083 ± 0.177 | − 0.264, 0.430 | 0.061 |
ACD (mm) | |||
Anterion vs. Argos | − 0.052 ± 0.076 | − 0.202,0.098 | 0.006* |
Anterion vs. IOLMaster 700 | 0.081 ± 0.029 | 0.024,0.138 | < 0.001* |
Anterion vs. Pentacam AXL | 0.026 ± 0.068 | − 0.107,0.159 | 0.004* |
Argos vs. IOLMaster 700 | 0.133 ± 0.078 | − 0.021,0.286 | < 0.001* |
Argos vs. Pentacam AXL | 0.078 ± 0.043 | − 0.007,0.163 | < 0.001* |
IOLMaster 700 vs. Pentacam AXL | − 0.055 ± 0.066 | − 0.183,0.074 | < 0.001* |
LT (mm) | |||
Anterio vs. Argos | 0.031 ± 0.075 | − 0.115,0.178 | < 0.001* |
Anterion vs. IOLMaster 700 | 0.048 ± 0.027 | − 0.006,0.102 | < 0.001* |
Argos vs. IOLMaster 700 | 0.023 ± 0.096 | − 0.166,0.211 | 0.005* |
Axial length (mm) | |||
Anterion vs. Argos | 0.013 ± 0.066 | − 0.116,0.142 | 0.998 |
Anterion vs. IOLMaster 700 | − 0.006 ± 0.033 | − 0.071,0.059 | 0.493 |
Anterion vs. Pentacam AXL | − 0.026 ± 0.181 | − 0.381,0.330 | 0.393 |
Anterion vs. OcuScan RxP | 0.146 ± 0.528 | − 0.888,1.181 | 0.148 |
Argos vs. IOLMaster 700 | − 0.028 ± 0.087 | − 0.199,0.143 | 0.701 |
Argos vs. Pentacam AXL | − 0.044 ± 0.204 | − 0.444,0.355 | 0.598 |
Argos vs. OcuScan RxP | 0.164 ± 0.515 | − 0.845,1.173 | 0.068 |
IOLMaster 700 vs. Pentacam AXL | − 0.023 ± 0.200 | − 0.415,0.369 | 1.000 |
IOLMaster 700 vs. OcuScan RxP | 0.180 ± 0.523 | − 0.845,1.205 | < 0.001* |
Pentacam AXL vs. OcuScan RxP | 0.234 ± 0.607 | − 0.950,1.428 | < 0.001* |