Skip to main content

Table 3 Inter-device agreement of corneal astigmatism in dioptric vector space

From: Comparison of corneal aberrations from anterior segment swept source OCT versus Placido-topography combined spectral domain OCT in cataract patients

Parameter

Mean ± SD

Mean difference ± SD

P

95% LoA

 

SS-OCT

Placido + SD-OCT

  

Lower, upper

Span

Anterior cornea (D)

      

 J0

 − 0.02 ± 0.39

 − 0.01 ± 0.38

 − 0.01 ± 0.25

0.71

 − 0.49, 0.51

1.00

 J45

0.11 ± 0.22

0.13 ± 0.21

0.02 ± 0.20

0.31

 − 0.37, 0.41

0.78

Total cornea (D)

      

 J0

0.13 ± 0.39

0.11 ± 0.36

 − 0.02 ± 0.18

0.38

 − 0.36, 0.33

0.69

 J45

0.11 ± 0.20

0.12 ± 0.20

0.02 ± 0.15

0.31

 − 0.29, 0.32

0.61

Posterior cornea (D)

      

 J0

0.14 ± 0.06

0.12 ± 0.15

 − 0.02 ± 0.15

0.11

 − 0.33, 0.28

0.61

 J45

0.00 ± 0.04

0.01 ± 0.06

 − 0.01 ± 0.05

0.20

 − 0.11, 0.09

0.20

  1. SD = standard deviation; D = diopter; LoA = limits of agreement; SD-OCT = spectral-domain optical coherence tomography; SS-OCT = swept-source optical coherence tomography