Skip to main content

Table 3 Agreement between the two devices for different parameters analyzed

From: Repeatability of a fully automated swept-source optical coherence tomography biometer and agreement with a low coherence reflectometry biometer

Parameter

Mean difference ± SD

95% CI

95% LoA

LoA interval

K1 (D)

 − 0.011 ± 0.100

 − 0.034, 0.011

 − 0.208, 0.185

0.393

K2 (D)

 − 0.024 ± 0.142

 − 0.056, 0.008

 − 0.302, 0.254

0.557

CCT (μm)

3.117 ± 4.026

2.199, 4.034

 − 4.774, 11.008

15.783

ACD (mm)

0.012 ± 0.054

0.000, 0.024

 − 0.093, 0.118

0.212

LT (mm)

 − 0.024 ± 0.088

 − 0.045, − 0.004

 − 0.198, 0.148

0.347

AL (mm)

0.004 ± 0.020

 − 0.001, 0.008

 − 0.036, 0.044

0.080

  1. K1 = flat keratometry; K2 = steep keratometry; CCT = central corneal thickness; ACD = anterior chamber depth; LT = lens thickness; AL = axial length; SD = standard deviation; CI = confidence interval; LoA = limits of agreement