Table 1 Repeatability analysis of the two optical biometers for the different parameters assessed

From: Repeatability of a fully automated swept-source optical coherence tomography biometer and agreement with a low coherence reflectometry biometer

Parameter/Biometer

Sw

CoV (%)

CoR

K1 (D)

   

 Eyestar 900

0.079

0.18

0.218

 Lenstar LS 900

0.073

0.17

0.202

K2 (D)

   

 Eyestar 900

0.075

0.17

0.209

 Lenstar LS 900

0.162

0.37

0.449

CCT (μm)

   

 Eyestar 900

1.559

0.29

4.320

 Lenstar LS 900

2.936

0.54

8.133

ACD (mm)

   

 Eyestar 900

0.018

0.61

0.050

 Lenstar LS 900

0.035

1.20

0.098

LT (mm)

   

 Eyestar 900

0.021

0.56

0.058

 Lenstar LS 900

0.066

1.74

0.184

AL (mm)

   

 Eyestar 900

0.008

0.03

0.022

 Lenstar LS 900

0.014

0.06

0.040

  1. K1 = flat keratometry; K2 = steep keratometry; CCT = central corneal thickness; ACD = anterior chamber depth; LT = lens thickness; AL = axial length; Sw = within subject standard deviation; CoV = coefficient of variability; CoR = coefficient of repeatability