Table 4 Repeatability of corneal thickness of all used devices

From: Comparison of corneal tomography using a novel swept-source optical coherence tomographer and rotating Scheimpflug system in normal and keratoconus eyes: repeatability and agreement analysis

Parameter

Device

Normal eyes

Mild KC

Moderate KC

Mean ± SD

Sw

Repeatability

CV (%)

Mean ± SD

Sw

Repeatability

CV (%)

Mean ± SD

Sw

Repeatability

CV (%)

CCT (µm)

RSC

556 ± 25

1.449

4.017

0.223

515 ± 31

2.366

6.559

0.307

466 ± 32

3.033

8.407

0.561

SS-OCT

549 ± 26

0.775

2.147

0.097

513 ± 30

0.775

2.147

0.134

462 ± 34

1.732

4.801

0.271

MCT (µm)

RSC

553 ± 25

1.761

4.880

0.272

501 ± 30

2.387

6.618

0.402

451 ± 31

3.033

8.407

0.566

SS-OCT

546 ± 27

0.775

2.147

0.095

493 ± 27

0.894

2.479

0.125

444 ± 30

1.225

3.395

0.196

  1. CCT = central corneal thickness; CV = coefficient of variation; KC = keratoconus; MCT = minimal corneal thickness; repeatability = coefficient of repeatability; RSC = rotating Scheimpflug camera; SS-OCT = swept-source optical coherence tomography; Sw = within-subject standard deviation