Skip to main content

Table 3 Repeatability of posterior corneal curvature parameters of all used devices

From: Comparison of corneal tomography using a novel swept-source optical coherence tomographer and rotating Scheimpflug system in normal and keratoconus eyes: repeatability and agreement analysis

Parameter

Device

Normal eyes

Mild KC

Moderate KC

Mean ± SD

Sw

Repeatability

CV (%)

Mean ± SD

Sw

Repeatability

CV (%)

Mean ± SD

Sw

Repeatability

CV (%)

Flat SimK (D)

RSC

 − 5.9 ± 0.3

0.035

0.098

− 0.362

 − 6.2 ± 0.4

0.105

0.291

 − 0.976

 − 7.2 ± 0.9

0.091

0.251

− 0.926

SS-OCT

 − 5.9 ± 0.2

0.005

0.015

− 0.078

 − 6.1 ± 0.4

0.017

0.048

 − 0.236

 − 7.3 ± 1.0

0.041

0.114

− 0.408

Steep SimK (D)

RSC

 − 6.3 ± 0.2

0.028

0.077

− 0.211

 − 6.8 ± 0.4

0.068

0.188

 − 0.722

 − 8.0 ± 1.0

0.074

0.206

− 0.734

SS-OCT

 − 6.2 ± 0.2

0.008

0.021

− 0.100

 − 6.9 ± 0.5

0.020

0.055

 − 0.258

 − 8.1 ± 1.0

0.033

0.092

− 0.312

Average SimK (D)

RSC

 − 6.1 ± 0.2

0.028

0.077

− 0.220

 − 6.5 ± 0.4

0.065

0.180

 − 0.726

 − 7.6 ± 0.9

0.064

0.177

− 0.681

SS-OCT

 − 6.0 ± 0.2

0.005

0.015

− 0.073

 − 6.5 ± 0.4

0.014

0.039

 − 0.190

 − 7.7 ± 1.0

0.033

0.092

− 0.340

Astigmatism (D)

RSC

 − 0.4 ± 0.2

0.036

0.101

− 7.107

 − 0.6 ± 0.3

0.120

0.333

 − 23.142

 − 0.8 ± 0.5

0.099

0.276

− 14.586

SS-OCT

 − 0.4 ± 0.2

0.009

0.025

− 2.720

 − 0.8 ± 0.4

0.022

0.062

 − 2.281

 − 0.8 ± 0.4

0.039

0.107

− 4.616

Kmax (D)

RSC

 − 6.5 ± 0.3

0.034

0.095

− 0.460

 − 8.2 ± 0.7

0.160

0.443

 − 1.570

 − 9.9 ± 1.5

0.176

0.488

− 1.459

SS-OCT

 − 6.3 ± 0.3

0.014

0.039

− 0.188

 − 8.1 ± 0.8

0.045

0.124

 − 0.407

 − 9.8 ± 1.5

0.071

0.196

− 0.519

BFS (mm)

RSC

6.6 ± 0.2

0.013

0.036

0.150

6.4 ± 0.2

0.016

0.045

0.213

6.1 ± 0.4

0.020

0.055

0.237

SS-OCT

6.7 ± 0.2

0.010

0.029

0.100

6.5 ± 0.2

0.008

0.023

0.099

6.1 ± 0.4

0.010

0.028

0.161

  1. BFS = best-fit sphere; CV = coefficient of variation; Kmax = maximum keratometry; KC = keratoconus; repeatability = coefficient of repeatability; RSC = rotating Scheimpflug camera; SimK = simulated keratometry value; SS-OCT = swept-source optical coherence tomography; Sw = within-subject standard deviation