Skip to main content

Table 4 Interobserver reproducibility for anterior corneal aberrations between the two observers

From: Repeatability and reproducibility of corneal higher-order aberrations measurements after small incision lenticule extraction using the Scheimpflug-Placido topographer

Parameter

Meanā€‰Ā±ā€‰SD (Ī¼m)

Sw (Ī¼m)

TRT (Ī¼m)

ICC (95% CI)

Total RMS

1.28ā€‰Ā±ā€‰0.42

0.04

0.12

0.989 (0.982ā€“0.993)

Higher-order RMS

0.64ā€‰Ā±ā€‰0.18

0.02

0.07

0.983 (0.973ā€“0.989)

Coma Z (3,ā€‰Ā±ā€‰1)

0.36ā€‰Ā±ā€‰0.20

0.03

0.07

0.984 (0.974ā€“0.990)

Trefoil Z (3,ā€‰Ā±ā€‰3)

0.18ā€‰Ā±ā€‰0.08

0.03

0.08

0.875 (0.802ā€“0.921)

SA Z (4, 0)

0.40ā€‰Ā±ā€‰0.12

0.01

0.03

0.989 (0.983ā€“0.993)

Ast II Z (4,ā€‰Ā±ā€‰2)

0.13ā€‰Ā±ā€‰0.09

0.01

0.04

0.974 (0.960ā€“0.984)

  1. SDā€‰=ā€‰standard deviation; Swā€‰=ā€‰within-subject standard deviation; TRTā€‰=ā€‰test-retest repeatability (2.77 Sw); ICCā€‰=ā€‰intraclass correlation coefficient; RMSā€‰=ā€‰root mean square; SAā€‰=ā€‰spherical aberration; Astā€‰=ā€‰astigmatism