Skip to main content

Table 3 Inter-observer reproducibility of ocular biological parameters measured using the new swept-source optical coherence tomography biometer in children before and after wearing orthokeratology lens

From: Effect of orthokeratology on precision and agreement assessment of a new swept-source optical coherence tomography biometer

Parameter

Status

Sw

R

CoV (%)

ICC

AL (mm)

Before

0.01

0.02

0.03

1.000

After

0.01

0.02

0.03

1.000

CCT (μm)

Before

2.36

6.53

0.44

0.994

After

2.53

7.02

0.48

0.994

ACD (mm)

Before

0.02

0.06

0.54

0.988

After

0.02

0.06

0.64

0.989

LT (mm)

Before

0.03

0.07

0.75

0.967

After

0.02

0.07

0.71

0.985

Km (Φ = 2.5) (D)

Before

0.06

0.17

0.14

0.998

After

0.08

0.21

0.19

0.998

Km (Φ = 3.0) (D)

Before

0.06

0.17

0.14

0.998

After

0.06

0.18

0.15

0.998

CD (mm)

Before

0.06

0.17

0.50

0.980

After

0.06

0.17

0.50

0.976

  1. AL= Axial length; CCT= central corneal thickness; ACD= anterior chamber depth; LT= lens thickness; Km= mean keratometry; CD= corneal diameter; SD= standard deviation; Sw= within-subject standard deviation; R= reproducibility limit (2.77 Sw); CoV= within-subject coefficient of variation; ICC= intraclass correlation coefficient