Table 5 Disparity between the recommended and implanted lens sphere values when targeting emmetropia

From: Artificial intelligence-based refractive error prediction and EVO-implantable collamer lens power calculation for myopia correction

 Parameters

MPE (D)

Range of PE (D)

SD (D)

MAE (D)

MedAE (D)

Percentage of eyes within the ranges (%)

   

 ± 0.25 D

 ± 0.5 D

 ± 0.75 D

NT-ICL lens sphere

0.016

 − 1.25 to 1.50

0.981

0.347

0.25

59.50

85.95

94.21

TICL lens sphere

 − 0.729a

3.25 to 2.50

1.420

1.221

1.00

23.98

33.16

45.41

  1. PE = prediction error; MPE = mean prediction error; MAE = mean absolute error, MedAE = median absolute error; SD = standard deviation; D = diopters
  2. The disparity between the recommended power and implanted lens power of NT-ICL did not differ significantly from zero, while that of TICL was lower with − 0.729 D than the implanted power. The recommended lens sphere deviated from the implanted lens sphere by < 0.25 D (one step provided by the manufacturer) in 56.92% of NT-ICL cases and < 0.50 D (one step provided by the manufacturer) in 32.20% of the time in TICL cases
  3. aStatistically significant difference from zero (P < 0.05)