Skip to main content

Table 2 Repeatability of anterior corneal curvature parameters of all used devices

From: Comparison of corneal tomography using a novel swept-source optical coherence tomographer and rotating Scheimpflug system in normal and keratoconus eyes: repeatability and agreement analysis

Parameter

Device

Normal eyes

Mild KC

Moderate KC

Mean ± SD

Sw

Repeatability

CV (%)

Mean ± SD

Sw

Repeatability

CV (%)

Mean ± SD

Sw

Repeatability

CV (%)

Flat SimK (D)

RSC

41.9 ± 1.1

0.071

0.196

0.129

43.2 ± 1.5

0.182

0.504

0.309

48.2 ± 4.8

0.283

0.784

0.414

SS-OCT

42.1 ± 1.1

0.095

0.263

0.184

43.4 ± 1.4

0.134

0.372

0.262

47.6 ± 4.6

0.173

0.480

0.305

Steep SimK (D)

RSC

43.2 ± 1.4

0.071

0.196

0.130

46.3 ± 2

0.187

0.519

0.271

52.0 ± 4.9

0.332

0.919

0.464

SS-OCT

43.4 ± 1.3

0.100

0.277

0.188

46.8 ± 2.0

0.145

0.402

0.271

51.9 ± 4.6

0.200

0.554

0.309

Average SimK (D)

RSC

42.5 ± 1.1

0.063

0.175

0.128

44.7 ± 1.6

0.145

0.402

0.248

50.0 ± 4.7

0.265

0.733

0.351

SS-OCT

42.7 ± 1.1

0.089

0.248

0.163

45.0 ± 1.5

0.114

0.316

0.217

49.6 ± 4.4

0.173

0.480

0.281

Astigmatism (D)

RSC

1.2 ± 1.0

0.069

0.190

7.456

3.2 ± 1.6

0.247

0.685

5.603

3.9 ± 2.1

0.313

0.868

10.038

SS-OCT

1.3 ± 1.0

0.082

0.229

8.618

3.4 ± 1.7

0.161

0.447

4.413

4.3 ± 2.2

0.192

0.533

4.600

Kmax (D)

RSC

43.7 ± 1.5

0.134

0.372

0.233

51.5 ± 3.3

0.224

0.620

0.369

59.7 ± 6.8

0.480

1.329

0.620

SS-OCT

43.8 ± 1.4

0.130

0.361

0.248

50.9 ± 3.1

0.179

0.496

0.308

58.1 ± 6.1

0.283

0.784

0.377

BFS (mm)

RSC

8.0 ± 0.2

0.009

0.025

0.071

7.7 ± 0.2

0.009

0.026

0.092

7.3 ± 0.4

0.014

0.039

0.144

SS-OCT

8.0 ± 0.2

0.010

0.028

0.099

7.6 ± 0.2

0.011

0.030

0.116

7.3 ± 0.4

0.010

0.028

0.120

  1. BFS = best-fit sphere; CV = coefficient of variation; Kmax = maximum keratometry; KC = keratoconus; repeatability = coefficient of repeatability; RSC = rotating Scheimpflug camera; SimK = simulated keratometry value; SS-OCT = swept-source optical coherence tomography; Sw = within-subject standard deviation