Table 5 Interobserver reproducibility for posterior corneal aberrations between the two observers

From: Repeatability and reproducibility of corneal higher-order aberrations measurements after small incision lenticule extraction using the Scheimpflug-Placido topographer

Parameter

Meanā€‰Ā±ā€‰SD (Ī¼m)

Sw (Ī¼m)

TRT (Ī¼m)

ICC (95% CI)

Total RMS

0.36ā€‰Ā±ā€‰0.11

0.02

0.05

0.976 (0.962ā€“0.985)

Higher-order RMS

0.11ā€‰Ā±ā€‰0.03

0.01

0.04

0.845 (0.766ā€“0.899)

Coma Z (3,ā€‰Ā±ā€‰1)

0.03ā€‰Ā±ā€‰0.02

0.00

0.01

0.967 (0.948ā€“0.979)

Trefoil Z (3,ā€‰Ā±ā€‰3)

0.06ā€‰Ā±ā€‰0.03

0.02

0.05

0.686 (0.544ā€“0.789)

SA Z (4, 0)

0.02ā€‰Ā±ā€‰0.02

0.01

0.01

0.886 (0.826ā€“0.926)

Ast II Z (4,ā€‰Ā±ā€‰2)

0.02ā€‰Ā±ā€‰0.01

0.00

0.01

0.819 (0.728ā€“0.882)

  1. SDā€‰=ā€‰standard deviation; Swā€‰=ā€‰within-subject standard deviation; TRTā€‰=ā€‰test-retest repeatability (2.77 Sw); ICCā€‰=ā€‰intraclass correlation coefficient; RMSā€‰=ā€‰root mean square; SAā€‰=ā€‰spherical aberration; Astā€‰=ā€‰astigmatism