Table 1 Patient demographics

From: Evaluation of rebound tonometer iCare IC200 as compared with IcarePRO and Goldmann applanation tonometer in patients with glaucoma

Variable Value
Eyes, R/L 122/23
Subjects, female/male 82/63
Age, mean ± SD (range), years 65.6 ± 11.6 (35 to 86)
Axial length, mean ± SD (range), mm 25.1 ± 1.8 (20.2 to 30.9)
Corneal curvature, mean ± SD (range), mm 7.61 ± 0.29 (6.89 to 8.45)
Central corneal thickness, mean ± SD (range), mm 0.522 ± 0.030 (0.412 to 0.606)
  1. SD standard deviation