Skip to main content

Table 3 Influence of MfIOL placement on vertical decentration among non-myopic and myopic eyes

From: Decentration and tilt of plate-haptic multifocal intraocular lenses in myopic eyes

Placement

Non-Myopic: Axial length ≤ 24.5 mm

Myopic: Axial length > 24.5 mm

Group A

(n = 30)

Group B

(n = 28)

P value*

Group A

(n = 33)

Group B

(n = 31)

P value*

Na

Vertical decentration (mm)

Na

Vertical decentration (mm)

Na

Vertical decentration (mm)

Na

Vertical decentration (mm)

Horizontal

7

−0.03 ± 0.06

6

− 0.13 ± 0.20

0.229

6

− 0.03 ± 0.10

9

−0.22 ± 0.11

0.005

Vertical

5

0.00 ± 0.06

7

−0.02 ± 0.20

0.823

6

−0.03 ± 0.07

6

−0.07 ± 0.13

0.537

Oblique

18

−0.03 ± 0.09

15

− 0.06 ± 0.09

0.289

21

−0.03 ± 0.10

16

−0.17 ± 0.14

0.001

  1. MfIOL = Multifocal intraocular lens
  2. Data are mean ± standard deviation
  3. * P values < 0.05 were considered statistically significant. Student’s t-test
  4. aN = Number of eyes