Skip to main content

Table 3 Influence of MfIOL placement on vertical decentration among non-myopic and myopic eyes

From: Decentration and tilt of plate-haptic multifocal intraocular lenses in myopic eyes

PlacementNon-Myopic: Axial length ≤ 24.5 mmMyopic: Axial length > 24.5 mm
Group A
(n = 30)
Group B
(n = 28)
P value*Group A
(n = 33)
Group B
(n = 31)
P value*
NaVertical decentration (mm)NaVertical decentration (mm)NaVertical decentration (mm)NaVertical decentration (mm)
Horizontal7−0.03 ± 0.066− 0.13 ± 0.200.2296− 0.03 ± 0.109−0.22 ± 0.110.005
Vertical50.00 ± 0.067−0.02 ± 0.200.8236−0.03 ± 0.076−0.07 ± 0.130.537
Oblique18−0.03 ± 0.0915− 0.06 ± 0.090.28921−0.03 ± 0.1016−0.17 ± 0.140.001
  1. MfIOL = Multifocal intraocular lens
  2. Data are mean ± standard deviation
  3. * P values < 0.05 were considered statistically significant. Student’s t-test
  4. aN = Number of eyes